§335.8230. Therapy-related Computer Systems for Remote Afterloader Units, Teletherapy Units and Gamma Stereotactic Units


Latest version.
  • The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies.  At a minimum, the acceptance testing must include, as applicable, verification of:

     

    a)         The source-specific input parameters required by the dose calculation algorithm;

     

    b)         The accuracy of dose, dwell time and treatment time calculations at representative points;

     

    c)         The accuracy of isodose plots and graphic displays;

     

    d)         The accuracy of the software used to determine sealed source positions from radiographic images; and

     

    e)         The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.

     