§335.7090. Therapy-related Computer Systems for Manual Brachytherapy  


Latest version.
  • The licensee shall:

     

    a)         Perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies.  At a minimum, the acceptance testing must include, as applicable, verification of:

     

    1)         The source-specific input parameters required by the dose calculation algorithm;

     

    2)         The accuracy of dose, dwell time, and treatment time calculations at representative points;

     

    3)         The accuracy of isodose plots and graphic displays; and

     

    4)         The accuracy of the software used to determine sealed source positions from radiographic images.

     

    b)         Maintain a record of acceptance testing and copies of the protocols used for acceptance testing in accordance with this Section for 5 years after discontinuation of use of the treatment planning system.

     

(Source:  Added at 30 Ill. Reg. 9029, effective April 28, 2006)